Lancet needle retention test
![Lancet retention test equipment on customer site Lancet retention test equipment on customer site](/sites/default/files/styles/case_study_gallery_540w/public/case-study/2024-04/lancet-retention-test-on-site.jpg.webp?itok=uXJwXW47)
![Lancet retention test fixtures on a MultiTest-dV Lancet retention test fixtures on a MultiTest-dV](/sites/default/files/styles/case_study_gallery_540w/public/case-study/2024-04/multitest-dv-lancet-fixture.jpg.webp?itok=ke8AExb0)
Solution
- Motor-driven tensile tester
- Needle lancet assembly supporting fixture
Benefits
- Quick results displayed immediately
- Optimum repeatbility with motor-drive
- Efficient QC testing at point of production
"We have been using Mecmesin equipment to perform retention tests on our range of lancet needles for over 8 years. We find the system easy to use for regular in-process checks, and generally reliable with prompt servicing from Mecmesin when required."
John Potter, Quality Systems Engineer
Owen Mumford
Requirement
A UK based medical device company who design and manufacture a range of blood glucose monitoring and control equipment used by healthcare professionals and consumers across the globe, needed a test solution to ensure the quality of their lancet needles. The company manufacture a variety of single use lancets, used by diabetes sufferers to obtain capillary blood samples when mounted into a lancing device.
The lancet comprises a steel needle moulded into a plastic base with a twist-off cap covering the point. To ensure the needle is not dislodged from its base when the cap is removed prior to use, stringent in-house production specifications dictate the minimum acceptable retention force of the needle within the plastic surround.
Solution
The company use a Mecmesin AFG 500 force gauge on an UltraTest stand (now superseded by the MultiTest 1) to perform retention tests on the lancet needles at regular 6-hourly intervals. This guarantees consistent manufacturing quality. The plastic base is placed under a custom-engineered mounting block, with the needle tip exposed vertically.
A lever-operated pin vice is fitted to the tip and a tensile load applied to the needle at a rate of 200 mm/min until it is dislodged from the plastic base. This durable Mecmesin system has been in constant use at the point of production for over 8 years.
Test equipment
- AFG 500 Force Gauge
UltraTest Stand (now superseded by the MultiTest 1-dV)
Lever-Operated Pin Vice
Custom-engineered mounting block (lower fixture)